Glow-Wire Testing to IEC 60695-2: GWEPT, GWFI and GWIT Explained
Overheated connections, failing components and overloaded contacts can turn a small part of an electrical product into a hot spot. The glow-wire test simulates that situation: a heated wire loop is pressed against the material or the finished part, and the laboratory observes whether it ignites, how long it burns and whether burning droplets could spread fire. It is required by many product standards, including IEC 60335-1 (household appliances), IEC 60598-1 (luminaires), IEC 60669-1 (switches), IEC 60884-1 (plugs and socket-outlets), IEC 60947-1 (low-voltage switchgear) and IEC 62368-1 (audio/video and ICT equipment).
The IEC 60695-2 series
The IEC 60695-2 series is split into four parts:
IEC 60695-2-10: the glow-wire apparatus and the common test procedure.
IEC 60695-2-11 (GWEPT): glow-wire flammability test for end products.
IEC 60695-2-12 (GWFI): glow-wire flammability index of materials.
IEC 60695-2-13 (GWIT): glow-wire ignition temperature of materials.
The same apparatus serves all three test methods. What changes is the specimen (finished product or standardised material plaque), the temperatures applied and the way the result is expressed.
How the test is performed
The specimen is mounted on a carriage and brought into contact with the glow-wire tip once the tip temperature has been stable within 5 K for at least 60 s. The approach speed is 10–25 mm/s, and contact must be gentle: the impact force must not exceed 1.05 N. The wire is held against the specimen for the application time Ta, typically 30 s, while the operator observes and records ignition (Ti), flame height and extinction (Te).
A specified layer, wrapping tissue on a wooden board, is placed below the specimen. If it ignites, burning droplets or particles from the specimen could spread fire in real use.
GWEPT, GWFI and GWIT: what each result means
GWEPT (end products): the finished part is tested at the glow-wire temperature required by its product standard. Typical values are 550, 650, 750, 850 or 960 °C, depending on the function of the part and whether it carries current. The result is pass or fail.
GWFI (materials): the highest temperature at which a material plaque of a given thickness meets the extinction criteria of the standard, including no ignition of the wrapping tissue. It describes how the material behaves once a flame appears.
GWIT (materials): 25 K (or 30 K) above the highest test temperature at which the material does not ignite, or does not flame continuously for more than 5 s in any single flame event. It describes how hard the material is to ignite.
For a designer, GWFI and GWIT values from the material supplier are a first screen. For certification, the product standard decides whether the material data are accepted or the end product must be tested.
Apparatus requirements
IEC 60695-2-10 specifies the apparatus in detail. The main requirements are:
Glow wire: NiCr (> 77 % Ni / 20 ± 1 % Cr), Ø 4.0 mm loop.
Thermocouple: type K, class 1, mineral-insulated and metal-sheathed, nominal Ø 1.0 mm in the current edition.
Contact force: 0.95 N ± 0.10 N (IEC 60695-2-10:2021 and 2026; earlier editions specified 1.0 N ± 0.2 N).
Penetration depth: 7.0 mm ± 0.5 mm, limited by a mechanical stop.
Test chamber: more than 0.5 m³, with a dark background (< 20 lx) so small flames are visible.
On the Ormity Glow Wire Test Apparatus, temperature is controlled from ambient to 1000 °C (test range 50–1050 °C) with ± 1 °C resolution and stability. The heating circuit uses a low-voltage, high-current transformer (0–170 A) with True RMS measurement. Timers record Ta (0–999.9 s), Ti and Te. Versions range from standard to motorized, automatic, and automatic with data acquisition. A version without hood, the GW1000, is also available.
Calibration at 960 °C
The temperature measuring system is checked with the silver-foil method: a piece of silver foil (purity at least 99.8 %, about 2 mm² and 0.06 mm thick) is placed on the tip, and when it starts to melt the indicated temperature must be 960 °C ± 10 °C. The 2026 edition also adds a normative annex on the use of a pyrometer. Ormity apparatus is supplied with silver foil for this check. If the indicated temperature at the melting point is outside tolerance, the thermocouple or its contact with the loop must be corrected before testing.
Common sources of error
Thermocouple drift. After many cycles at high temperature the thermocouple ages. Check it regularly with silver foil and replace it when needed; see the glow-wire heating element.
Wrong contact force. A carriage that sticks or a mis-set counterweight changes the penetration and the heat transfer.
Light in the chamber. Small, short flames are easily missed in a bright room.
Wrong indicator layer. Use the specified wrapping tissue and board. Our glow-wire indicator cloth is listed in the consumables range.
Specimen conditioning. Follow the conditioning required by the standard before testing.
Upgrades for existing apparatus
Clamping jaws heat up during long test series. The water-cooled glow-wire clamping jaws retrofit kit is designed for existing glow-wire apparatus.
For laboratories: Technical Note 03: Glow-Wire Apparatus Verification and GWFI / GWIT Test Planning gives the full apparatus verification and GWFI / GWIT test sequences.
FAQ
Is UL 94 the same as the glow-wire test?
No. UL 94 uses a flame applied to a material specimen. The glow-wire test uses a heated wire without an open flame. Product standards may refer to both, and UL 746A lists glow-wire results among the properties of polymeric materials.
Which temperature should I test at?
For end products, the temperature is set by your product standard. For materials (GWFI and GWIT), the test determines the temperature.
How often should the apparatus be calibrated?
Check the temperature with silver foil at regular intervals defined by your quality system, and after replacing the thermocouple or the loop. Periodic calibration of the measuring chain by an ISO/IEC 17025 accredited laboratory gives traceable results; see our ISO/IEC 17025 thermocouple calibration service.