Automatic CTI tracking test apparatus to IEC 60112

CTI and PTI Testing to IEC 60112: How a Tracking Test Apparatus Works

Nov 3
3 min read

When an insulating surface is wet and contaminated, small leakage currents can flow between conductors. Over time they carbonise the surface and form a conductive path: this is tracking. The comparative tracking index (CTI) measures how well a solid insulating material resists it. Designers use the CTI to choose materials and to determine creepage distances, which is why the value appears in material datasheets, in insulation coordination tables and in certification files.

CTI and PTI

IEC 60112 defines two indices:

  • CTI (comparative tracking index): the highest voltage the material withstands under the specified test conditions without tracking. It is used to classify and compare materials.

  • PTI (proof tracking index): a pass/fail check that the material withstands a given voltage. It is used to verify that a material meets a specified level.

The same apparatus and the same procedure are used for both; only the way the test voltage is chosen and the result reported differ. ASTM D3638 describes a comparable comparative tracking index method, and UL 746A uses CTI results to classify polymeric materials.

How the test is performed

Two electrodes are placed on the specimen surface under a defined force. Drops of contaminant solution fall between them at fixed intervals while the selected test voltage is applied. If the material withstands the specified number of drops without tracking, it passes at that voltage. Repeating the test at different voltages gives the CTI; a single test at a specified voltage gives the PTI. A failure is a current of 0.5 A or more flowing for 2 s, or persistent flaming. Edition 5.0 also introduced a screening test for materials that withstand high test voltages.

The reference contaminant is solution A, ammonium chloride in water. Edition 5.0 (2020) added solution C, with a surfactant aligned with IEC 60587, and moved solution B to an annex for continuity with earlier results. The solution used must be stated with the result; results obtained with solution B carry the suffix M.

Apparatus

The accuracy of a CTI result depends on the electrodes, the drops and the electrical circuit. On the Ormity CTI Tracking Test Apparatus:

  • Electrodes: platinum, 5 × 2 mm, 30° chisel, spacing 4.0 mm, force 1.00 N.

  • Test voltage: 100–600 V AC.

  • Dropping: electromechanical pipetting of solution A or B at 30 s intervals.

  • Short-circuit check: automatic 1 A circuit verification before each test.

  • Resistor bank: automatic switching.

  • Operation: fully automatic, with a 10" colour touch-screen HMI.

  • Safety: electrically assisted door with safety glass; effluent collection.

Two measuring versions are available: a conventional measuring chain or electronic True-RMS measurement.

Typical sources of error

  • Electrode condition. The chisel edges must be clean and correctly shaped, and the spacing exactly 4.0 mm. Replacement platinum CTI electrodes are available.

  • Drop size and timing. Manual dropping is a major source of scatter. An automatic electrolyte dispenser delivers consistent drops.

  • Specimen surface. Surfaces must be clean and flat; follow the preparation and conditioning required by the standard.

  • Circuit verification. The short-circuit current must be verified before testing, which the apparatus does automatically.

IEC 60112 or IEC 60587?

IEC 60112 is a laboratory test on small specimens at low voltage (up to 600 V), using drops of contaminant. IEC 60587 evaluates materials for use under severe ambient conditions at higher voltages, with a continuous flow of contaminant on an inclined specimen. They answer different questions and are not interchangeable. For IEC 60587, see our inclined plane tracking test apparatus and the hybrid AC/DC tracking and erosion test apparatus.

CTI simulator

Our free CTI tracking index simulator shows the test interactively and can be used for operator training.

For laboratories: Technical Note 05: Reliable CTI and PTI Results to IEC 60112 gives the laboratory checks, step sequence and reporting rules.

FAQ

Is a higher CTI always better?

A higher CTI means better resistance to tracking under the test conditions, which may allow shorter creepage distances. Other properties, such as glow-wire behaviour and mechanical strength, still need to be checked.

Can I measure CTI above 600 V?

IEC 60112 covers test voltages up to 600 V. Materials for higher voltages and severe environments are evaluated with IEC 60587.

What is the difference between solutions A, B and C?

Solution A is ammonium chloride in water. Solution B adds a wetting agent and is more aggressive; in IEC 60112:2020 it is kept in an annex. Solution C, introduced in 2020, uses a surfactant aligned with IEC 60587. The solution used must be reported with the result.

How do I keep the results traceable?

Keep records of electrode inspection, drop calibration and the electrical circuit. Periodic calibration of the measuring chain by an ISO/IEC 17025 accredited laboratory gives traceable results.