Dosificador de electrolito CTI – IEC 60112
SKU ORTYTK112-50LSI
Precio
4490,00 CAD
CTI drop dispenser with motor-actuated electronic pipette for tracking tests, per IEC 60112.
Normas
- IEC 60112 – CTI / PTI drop application
Datos técnicos
| Nominal drop volume | 20 µL (verification 25 µL) |
| Drops per test | 50 (≈ 1.0 mL) |
| Drop interval | 30 s ± 5 s |
| Actuator | Micro-stepping motor NEMA 12, PTFE tip |
| Pipette required | Single-channel electronic, 5–50 µL, resolution ≤ 0.1 µL, calibrated |
| Systematic / random error at 25 µL | ± 0.25 µL (± 1 %) each, pipette-dependent |
| Environment | 20 ± 5 °C, ≤ 75 % RH |
| Column | Rigid aluminum |
Ref.: DNTK112-50LSI (Dunel) / ORTYTK112-50LSI (Ormity) – former codes: OTY112-500-SLI-Y
Cantidad
Principio de ensayo y funcionamiento
Mechanical actuation module for Dunel / Ormity CTI apparatus. Volumetric performance is defined by the electronic pipette (not supplied unless specified), which is calibrated separately (recommended every 12 months, ISO/IEC 17025, ILAC MRA traceable).
Pedido, calibración y garantía
- Plazo de entrega: hasta 35 días con calibración de fábrica Ormity; hasta 60 días con informe de calibración ISO/IEC 17025 (facturado aparte).
- Garantía: 1 año, más garantía de por vida contra la no conformidad (excepto revisiones de normas publicadas después de la fecha de venta).
- Pago: los pedidos internacionales se pagan al 100 % por adelantado. Todas las ventas son definitivas después de 30 días.
- Envío: mediante transportistas internacionales externos; el tiempo de tránsito depende del destino, del embalaje (caja/palé) y de la ruta de calibración. Solicite una estimación de entrega antes de pedir.