top of page
Logo Ormity
Abstract diagonal blue stripes background

PRODUIT

Accueil   /   Produit

Test Talc < 75 µm – IEC 60529 IP5X / IP6X – 10 kg

SKU ORM-TLC-75-B10
Prix

0,00 $

Asbestos-free test talc (hydrated magnesium silicate), 100% passing a 75 µm sieve, for the IP5X and IP6X dust tests of IEC 60529 and Test La (non-abrasive fine dust) of IEC 60068-2-68. Verified and certified lot by lot in Montréal, Canada.

 

Standards

  • IEC 60529 – 13.4 IP5X and 13.5 IP6X (talc, 2 kg/m³ of chamber volume, no more than 20 tests per charge)
  • IEC 60068-2-68 – Test La, non-abrasive fine dust (talc)
  • Product standards that refer to the IEC 60529 dust test for IP5X / IP6X codes

 

Technical data

MaterialTalc, Mg3Si4O10(OH)2 – CAS 14807-96-6
Particle size100% passing 75 µm (square-mesh sieve, 75 µm aperture, 50 µm wire – IEC 60529 13.4, ISO 3310-1)
Moisture< 0.5% (loss on drying at 105 °C)
AsbestosNot detected (verified per lot)
Typical compositionSiO2 58–63%, MgO 29–32%, LOI ≤ 6%
AppearanceFine white powder, odourless
Packaging5 US gal (18.9 L) plastic pail with lid and metal handle – nominal net weight 10 kg – bilingual EN/FR label
QualityCertificate of Analysis (CoA) per lot: moisture, 75 µm sieve result, SiO2, MgO, LOI, asbestos-free confirmation

 

Ref.: ORM-TLC-75-B10

Quantité

Use & Storage

Keep the pail tightly closed in a dry, covered place. Damp talc agglomerates and no longer passes the sieve: once opened, check moisture and sieving before testing.

 

Not intended for tests that require ISO 12103-1 test dust (e.g. ISO 20653) or for abrasive sand and dust tests (MIL-STD-810H Method 510, RTCA DO-160G Section 12).

Quality Control & Traceability

  • Sieve verification in Montréal with a certified test sieve (ISO 3310-1, 75 µm aperture, 50 µm wire) before lot release.
  • Each pail carries the product code, ORMITY lot number and fill date; the CoA ships with every delivery.
  • Full particle size distribution by laser diffraction at an outside laboratory, on request.
  • Certified 75 µm go/no-go sieve available for incoming inspection.

Safety

Read the Safety Data Sheet (SDS, WHMIS 2015) before use. Avoid generating and breathing dust: use local exhaust ventilation, suitable respiratory protection, safety glasses and gloves. Non-flammable material.

Aucun avis pour le momentPartagez votre expérience, soyez le premier à laisser un avis.
Fiche technique

Articles similaires

bottom of page